Laser epilator – Epila

Serious risk Damage to sight, Electric shock, Injuries

Published by Safety Gate 8 Mar 2007In the weekly report of 9 Mar 2007 (week 9, 2007)Safety Gate noticeJSON
latest report: week 43, 30 Oct 2026checked 1 min agoSource: EU Safety Gate
Product
Laser epilator
Brand
Epila
Brand as notified
EPILA
Type / model
Laser SI-808
Risk
This product poses a serious risk of damage to sight, skin lesions and electric shock because of insufficient insulation and protection against laser radiation if used incorrectly. In particular:- the laser beam is highly effective at a distance of 2-3 cm from the spacer on all settings (low, middle, high). This is sufficient to cause irreversible retina damage, among other things;- it is likely that the product can be classified as Laser Class 4 contrary to Class IIIb as labelled; - the product has accessible live parts through the power supply voltage to earth of 230 V~/50Hz causing risk of electric shock; - the product does not have:the required access control (keypad code, key, etc.) or the required time switch that reacts when the device is idle and automatically switches it off; - the required safety lock (the laser can be operated without any controlled function enable, e.g. body resistance measurement, or similar).Moreover, even if the instructions recommend that the product device be kept out of children's reach, the possibility of damage resulting from foreseeable misuse cannot be ruled out, mainly due to lack of access control and safety lock. The product does not comply with the Low Voltage Directive and the relevant requirements of the European standard EN 60335.
Measures
  • Voluntary withdrawal from the market and recall from consumers
Notifying country
Germany
Country of origin
Republic of Korea
Alert type
Consumer
Photos
Photo 1 of the product in Safety Gate alert 0216/07Photo 2 of the product in Safety Gate alert 0216/07
Photos shown from the Safety Gate server (European Commission); not copied here.

This page reproduces the Safety Gate record as published. recallarchive is not affiliated with the European Commission. If you own this product, follow the measures above and contact the seller.

Sources

  1. Safety Gate alert 0216/07
    European Commission, Safety Gaterecord fields marked “official alert record”: © European Union, CC BY 4.0; values listed separately here
  2. Safety Gate weekly report Report-2007-9 (XML)
    the copy this page was built fromCC0 1.0