Skip to content

Skin treatment device – JMF “Dermapen”

Serious risk Electric shock

Published by Safety Gate 3 Mar 2015In the weekly report of 13 Mar 2015 (week 10, 2015)Safety Gate noticeJSON
latest report: week 43, 30 Oct 2026checked just nowSource: EU Safety Gate
Product
Skin treatment device
Brand
JMF
Name
Dermapen
Type / model
HJ120100
Batch
Unknown
Description
Hand held pen style appliance with many tiny needles for skin treatment, supplied with a mains transformer and a control box, accompanied by a plug in adaptor to convert the fitted 2 pin plug to a UK 3 pin configuration. Packaged in plain white cardboard box.
Risk
null The power supply does not have adequate insulation and is not designed to withstand foreseeable overloads. The user may consequently receive a mains voltage electric shock. The lack of shutters on the plug adaptor may result in the user receiving an electric shock when plugging / unplugging the appliance. The product does not comply with the requirements of the Low Voltage Directive and the relevant European standard EN 61558 and the relevant national standard.
Measures
  • Destruction of the product · from 26 Feb 2015
Notifying country
United Kingdom
Country of origin
People's Republic of China
Alert type
Consumer
Photos
Photo 1 of the product in Safety Gate alert A12/0321/15Photo 2 of the product in Safety Gate alert A12/0321/15
Photos shown from the Safety Gate server (European Commission); not copied here.

This page reproduces the Safety Gate record as published. recallarchive is not affiliated with the European Commission. If you own this product, follow the measures above and contact the seller.

Sources

  1. Safety Gate alert A12/0321/15
    European Commission, Safety Gaterecord fields marked “official alert record”: © European Union, CC BY 4.0; values listed separately here